iol

AUROLAB’S
AUROVUE

FIRST PIECE
FOLDABLE IOL

aurovue

Technical specifications

Design Characteristics
Optic Diameter6.0mm
Overall Length12.5mm
Haptics DesignForce Enduring haptics
ConvexityBiconvex
Angulation3° Angulation
A Constant118.4 (Ultra Sound), 118.7 (Optical)
ACD5mm
Diopter Range10.0 to 15.0 in 1.0 D
15.0 to 30.0 in 0.5 D
Delivery SystemDisposable Injector and Cartridge
AC28-D1100 2.2mm wound assisted
Material Characteristics
Model NumberHP7600
Lens MaterialHydrophobic acrylic
10% UV Cut off385 nm
Light TransmittanceMore than 90%
Glass Transition Temperature8°C
Refractive Index1.47

Aurolab's Aurovue is the first indigenously developed hydrophobic single piece foldable IOL. It has been manufactured to combine the benefits offered by the material and an intelligently engineered design, suitable to most eyes.

  • Excellent Bio Compatibility

  • Aurovue material was developed in collaboration with technical experts from various countries. This material was subject to invitro and invivo studies and is proven to be highly bio compatible.

  • Excellent Mechanical Properties

  • Aurovue has been tested for mechanical properties such as compression force, dynamic fatigue, optic tilt etc. as required by ISO 11979. Results of these tests acknowledge the good design of the lens hence it can be implanted through 2.8 mm incision.

  • Effectively Reduces PCO

  • - Aurovue's material is tacky in nature that adheres well to the capsular bag after implantation

    - Aurovue has 3° vault at optic-haptic junction

    These two features of the IOL enhances adherence of the lens with the capsular bag and greatly inhibits the formation of PCO.

  • Low Glass Transition Temperature

  • Aurovue's Tg is at 8°C which is lesser keeps the material flexible even at low temperatures.

  • Higher Abbe Number: 55

  • Aurovue has a higher Abbe number of 55 which reduces chromatic aberration. Hence provides better color perception to the patient.

aurovue

Force Enduring Haptics, Greater Stability

The unique feature in Aurovue is that the haptics are thicker than the optic edge. The thickness, combined with the design, makes the haptics 'Force-Enduring'. Together, they enable the haptics to sustain/ absorb the pressure of capsular retraction and thus offers total stability to the optic. The chances of decentration are minimized to a great extent with these 'Force- enduring' haptics.

Adaptability to
Different capsular Sizes

The optic and haptics have been suitably designed to fit in the capsular bag of varied sizes. The lens adapts well to the postoperative retraction of the capsular bag and ensures centration and stability of the lens. The angle of contact made by the haptics with the wall of capsular bag is greater with Aurovue.

evgold

Optimal Refractive index, Less Refelction

This material has an optimal refractive index of 1.47. This offers better vision by doing away with the problem of greater light reflectivity posed by materials with higher refractive index.

Memory in the Capsular Bag

The lens unfolds slowly after implantation in the eye. This aids the surgeon to manipulate the lens in the bag easily. Also, the gradual unfolding enables easier removal of viscoelastic fluids from the capsular bag.

aurovue

Comes with a disposable
delivery system

Aurovue is the first hydrophobic lens that is made available with the disposable injector and cartridge. Implantation is made absolutely hassle free and the lens can be implanted through a 2.8 mm incision.